丙烯腈检测管128SD ACRYLONITRILE CH2=CHCN 0.2-20ppm
Tube 28SD
1. PERFORMANCE
1) Measuring range :1-20 ppm 0.5-10 ppm 0.25-5 ppm 0.2-4 ppm
Number of pump strokes 1(100mr) 2(200mr) 4(400mr) 5(500mr)
2) Sampling time :1.5 minutes/1 pump stroke
3) Detectable limit :0.05ppm(500mr)
4) Shelf life :1 year (Necessary to store in refrigerated place ; 0~10℃)
5) Operating temperature :0~40℃
6) Temperature compensation :Necessary(See "TEMPERATURE CORRECTION TABLE")
7) Reading :Direct reading from the scale calibrated by 1 pump stroke
8) Colour change :Yellow→Red
2. RELATIVE STANDARD DEVIATION
RSD-low:10% RSD-mid.:5% RSD-high:5%
3. CHEMICAL REACTION
By decomposing with an Oxidizer, Hydrogen cyanide is produced, Hydrogen cyanide reacts with Mercuric
chloride, then Hydrogen chloride is liberated and PH indicator is discoloured.
CH2=CHCN+CrO3+H2SO4→HCN
HCN+HgCI2→HCI
4. CALIBRATION OF THE TUBE
GAS CHROMATOGRAPHY
5. INTERFERENCE AND CROSS SENSITIVITY
125
15
20
10
ppm
100ml
GAS To pump
(Reddish brown)
PRETREAT TUBE DETECTOR TUBE
128SD ACRYLONITRILE
Substance Interference ppm Coexistence
Butadiene
Less than
350 ppm Not affected.
Toluene
Less than
600 ppm 〃
Hexane
Less than
800 ppm 〃
Styrene
Less than
720 ppm 〃
(NOTE)
In case of 2 to 5 pump strokes can be used to detemine the lower concentration.
Following formula is available for actual concentration.
Actual concentration=Tmperature corrected value×
TEMPERATURE CORRECTION TABLE
Scale True Concentration (ppm)
Readings 0℃ 10℃ 20℃ 30℃ 40℃
(ppm) (32°F) (50°F) (68°F) (86°F) (104°F)
20 26 22.5 20 18 17
18 23 20 18 16.5 15.5
16 20.5 18 16 15 14
14 18 15.5 14 13 12
12 15.5 13.5 12 11 10.5
10 13 11 10 9.5 9
8 10.5 9 8 7.5 7
6 8 7 6 6 5.5
4 5.5 4.5 4 4 4
2 3 2 2 2 2
1 1.5 1 1 1 1
1