乙酸乙酯检测管ETHYL ACETATE CH2=C(CH3)CO2C4H9 20-1000ppm

111U乙酸乙酯检测管ETHYL ACETATE CH2=C(CH3)CO2C4H9 20-1000ppm

参考价: 面议

具体成交价以合同协议为准
2022-06-02 18:20:02
175
产品属性
关闭
上海境宣实业有限公司

上海境宣实业有限公司

初级会员1
收藏

组合推荐相似产品

产品简介

甲基丙烯酸 111U ETHYL ACETATE (+) 乙酸乙酯,乙酸乙酯气体检测管ETHYL ACETATE CH2=C(CH3)CO2C4H9 20-1000ppm,检测管内的检测剂有效成份含量的大小与被检测气体的浓度成正比例。载体上所吸附的试剂量决定该检测管的量程范围。为追求更小的检测量而使试剂减少到更小,会造成色阶颜色变化浅淡,而不宜观察。

详细介绍

乙酸乙酯检测管111U ETHYL ACETATE CH2=C(CH3)CO2C4H9 20-1000ppm

Tube 11U
1. PERFORMANCE
1) Measuring range :10-1,000 ppm
Number of pump strokes 1(100mL)
2) Sampling time :1.5 minutes / 1 pump stroke
3) Detectable limit : 5 ppm
4) Shelf life :2 years
5) Operating temperature :10~40℃
6) Temperature compensation :Necessary (See "TEMPERATURE CORRECTION TABLE" )
7) Reading :Direct reading from the scale calibrated by 1 pump stroke
8) Colour change :Yellow→Brown
2. RELATIVE STANDARD DEVIATION
RSD-low :10 % RSD-mid.:10 % RSD-high :5%
3. CHEMICAL REACTION
Chromium oxide is reduced.
CH 3CO 2C 2H 5+Cr 6++H 2SO 4→Cr 3+
4. CALIBRATION OF THE TUBE
GAS CHROMATOGRAPHY
5. INTERFERENCE AND CROSS SENSITIVITY
GAS To pump
111U ETHYL ACETATE
TEMPERATURE CORRECTION COEFFICIENT TABLE
Scale Correction Coefficient (at 20℃)
Readings 10 ℃ 15 ℃ 20 ℃ 25 ℃ 30 ℃ 35 ℃ 40℃
(ppm) (50 °F) (59°F) (68°F) (77 °F) (86°F) (95°F) (104°F )
1,000 1.33 1.17 1.0 0.87 0.74 0.64 0.53
Substance Interference Coexistence
Alcohols Similar stain is produced. Higher reading are given.
Esters 〃〃
Ketones 〃〃
Aromatic hydrocarbons 〃〃
Whole reagent is discoloured
It the top of brown stain is
Paraffin hydrocarbons clear, the accuracy of
Pale brown.
readings is not affected.
Halogenated hydrocarbons
〃〃
FIG.1
0
200
400
600
800
1000
Trichloroethylene 〔ppm〕
FIG.1 Influence of Trichloroethylene
E thyl acetate indicating value 〔ppm 〕
100 200 300
800 1.38 1.19 1.0 0.86 0.73 0.63 0.53
600 1.40 1.20 1.0 0.86 0.72 0.63 0.53
400 1.40 1.20 1.0 0.85 0.70 0.58 0.46
200 1.40 1.20 1.0 0.84 0.68 0.55 0.42
100 1.50 1.25 1.0 0.81 0.62 0.48 0.33
50 1.50 1.25 1.0 0.77 0.54 0.43 0.32
30 1.50 1.25 1.0 0.77 0.53 0.42 0.30
10 1.50 1.25 1.0 0.75 0.50 0.40 0.30
www.r-biopharm.com.cn
www.r-biopharm.com.cn

上一篇:氧气的浓度决定了危险程度 下一篇:怎么选到合适的气体检测仪?
提示