品牌
代理商厂商性质
上海市所在地
二甲醚CH3OCH3气体检测管123S DIMETHYL ETHER 0.01-1.2%
Tube No.
1. PERFORMANCE
1) Measuring range :0.01-1.2%
Number of pump strokes 1(100ml)
2) Sampling time :1.5 minutes/1 pump stroke
3) Detectable limit :10 ppm
4) Shelf life :3 years
5) Operating temperature :0~40℃
6) Reading :Direct reading from the scale calibrated by 1 pump stroke
7) Colour change :Orange→Dark brown
2. RELATIVE STANDARD DEVIATION
RSD-low:10% RSD-mid.:10% RSD-high:10%
3. CHEMICAL REACTION
Potassium dichromate is reduced.
CH3OCH3+Cr6++H2SO4→Cr3+
4. CALIBRATION OF THE TUBE
GAS CHROMATOGRAPHY
5. INTERFERENCE AND CROSS SENSITIVITY
GAS To pump
1.2
1.0
0.8
0.6
0.4
0.2
0.1
0.05
0.01
%
123S DIMETHYL ETHER
Substance Interference ppm Coexistence
Acetylene 3% Whole reagent is changed to
Brown.
Propane 0.2% 〃
Oexthceepr to Hrgaalongice gnaasteeds hoyr dvraopcoaurbrsons Similar stain is produced. Higher readings are given.