品牌
经销商厂商性质
苏州市所在地
7440安规综合分析仪特点:
交/直流耐压、绝缘阻抗、接地阻抗测试
20×2 点矩阵 LCD 显示器
可选配内建4组或8组高压矩阵式扫描信道(7420, 7430 & 7440)
内建50组程序记忆、每一组8个测试步骤可进行程序记忆步骤的连结测试
测试线路阻抗自我归零校整功能,确保测试结果的准确性
内建PLC Remote遥控界面
键盘锁定功能
可外接标准GPIB/RS232/Print Card界面作控制
全范围电压输出的稳压率在1%以内
7440安规综合分析仪MODEL | 7410 | 7420 | 7430 | 7440 |
AC WITHSTAND VOLTAGE | ||||
Output Rating | 5KVAC/30mA | 5KVAC/30mA | 5KVAC/40mA | 5KVAC/40mA |
Range | Resolution | Accuracy | ||
Output Voltage,ACV | 0 - 5000 | 1 | ±(2% of setting + 5V) | |
Output Frequency | 50Hz/60Hz ±100ppm, User Selection | |||
Output Waveform | Sine Wave ,THD.<2% (Resistive Load), Crest Factor = 1.3 - 1.5 | |||
Output Regulation | ± (1% of output + 5V), From no load to full load | |||
SETTINGS | ||||
Hi-Limit AC Current, mA | 0-30.00 | 0-30.00 | 0-40.00 | 0-40.00 |
Resolution : 0.01, Accuracy : ±(2% of setting + 3counts) | ||||
Lo-Limit AC Current, mA | Range : 0 - 9.999 | |||
Resolution : 0.001, Accuracy : ±(2% of setting + 3counts) | ||||
Ramp Time, second | 0.1 - 999.9 | 0.1 | ±(0.1% + 0.05sec) | |
Dwell Time, second | 0, 0.3 - 999.9 (0=continuous) | 0.1 | ||
Arc Detection | 1 - 9 ranges (9 is the most sensitivity) | |||
AC Current Offset, mA | 0 - 2.000 or Auto Set | |||
7440DC WITHSTAND VOLTAGE | ||||
Output Rating | 6KVDC/9999µA | |||
Output Voltage, DCV | 0 - 6000 | 1 | ±(2% of setting + 5V) | |
Output Ripple | <5%(6KV/9999µA at Resistive Load) | |||
SETTINGS | ||||
Hi-Limit DC Current, µA | 0 - 9999 | 1 | ±(2% of setting + 3counts) | |
Lo-Limit DC Current, µA | 0 - 999.9 | 0.1 | ±(2% of setting + 3counts) | |
Ramp Time, second | 0.1 - 999.9 | 0.1 | ±(0.1% + 0.05sec) | |
Dwell Time, second | 0, 0.4 - 999.9 (0=continuous) | 0.1 | ||
Arc Detection | 1 - 9 ranges (9 is the most sensitivity) | |||
DC Current Offset, ?A | 0 - 200.0 or Auto Set | |||
Ramp High Current | ON/OFF, User Selection | |||
Charge - Low Current, µA | 0 - 350.0 or Auto Set | |||
Discharge Time | ≦ 200 msec | |||
7440INSULATION RESISTANCE | ||||
Output Rating | 1KVDC/9999MΩ | |||
Output Voltage, DCV | 100 - 1000 | 1 | ±(2% of setting + 3V) | |
Hi-Limit Resistance, MΩ | 0, 1 - 9999 (0=OFF) | 1 | ±(2% of setting+2counts)at 500-1000VDC, 1 -999.9 MΩ ±(8% of setting+2counts)at 500-1000VDC, 1000 -9999 MΩ ±(8% of setting+2counts)at 100-499VDC, 1 -999.9 MΩ | |
Lo-Limit Resistance, MΩ | 1 - 9999 | 1 | ||
Delay Time, second | 0, 0.5 - 999.9 (0=continuous) | 0.1 | ±(0.1% + 0.05sec) | |
Charge - Low Current, µA | 0 - 3.500 or Auto Set | |||
7440GROUND BOND (7440 ONLY) | ||||
Output Rating | 30A/600mΩ, 8V | |||
Output AC Current, A | 3.00 - 30.00 | 0.01 | ± (2% of setting +3 counts) | |
Output AC Voltage, V | 3.00 - 8.00 | 0.01 | ± (2% of setting +3 counts) at Open circuit | |
Output Frequency | 50Hz/60Hz ±100ppm, User Selection | |||
SETTINGS | ||||
Hi-Limit Resistance, mΩ | 0 - 150 (3 - 30A) | 1 | ± (2% of setting +2 counts) | |
Lo-Limit Resistance, mΩ | 0 - 600 (3 - 10A) | |||
Lead Resistance Offset, mΩ | 0 -200 | 1 | ± (2% of setting +2 counts) | |
Dwell Time, second | 0, 0.5 - 999.9 sec (0=continuous) | 0.1 | ±(0.1% + 0.05sec) | |
7440MEASUREMENT | ||||
AC/DC Voltage, KV | 0 - 6.00 | 0.01 | ±(1.5% of reading + 1count) | |
DC Voltage, V (IR) | 100 - 1000 | 1 | ±(1.5% of reading + 3V) | |
AC Current Range, mA | 0-3.500 | 0-3.500 | 0-3.500 | 0-3.500 |
3.00-30.00 | 3.00-30.00 | 3.00-40.00 | 3.00-40.00 | |
AC Current Resolution, mA | 0.001/0.01 | |||
AC Current Accuracy, mA | ± (2% of reading + 3counts) | |||
DC Current Range, µA | 0-350.0 | 0.1 | ± (2% of reading + 3counts) | |
300 - 3500 | 1 | |||
3000-9999 | 10 | |||
AC Current, A (GB) | 0 - 35.00 | 0.01 | ±(2% of reading+3counts) | |
Resistance, , MΩ(IR) | 1 - 9999 (Auto Range) | 0.001 0.01 0.1 1 | ±(8% of reading+2counts)at 100-499VDC, 1-999.9 MΩ ±(8% of reading+2counts)at 500-1000VDC, 1000 -9999 MΩ ±(2% of reading+2counts)at 500-1000VDC, 1 -999.9 MΩ | |
Resistance, , mΩ (GB) | 0 - 600 | 1 | ±(2% of reading+2counts) | |
7440GENERAL | ||||
Input Voltage AC | 115/230Vac±15%, 50/60Hz ± 5%, Fuse 5A Slow for 7410/7412 model, 6.3A Slow for 7420/7430/7440 model | |||
PLC Remote Control | Input : Test , Reset , Memory1.2.3 | |||
Output : Pass , Fail , Processing | ||||
Memory | 50 memories, 8 steps/memory | |||
Display | 20×2 LCD with back light | |||
Key Lock | To prevent unauthorized alteration of the test Parameters | |||
Calibration | Build-in software and external calibrated meters | |||
Alarm Volume Setting | Range: 0-9 ;0=OFF, 1 is softest volume, 9 is loudest volume. | |||
Rear Panel Output | YES | YES | YES | YES |
Built-in Scanner | - | 8channels | Option | Option |
External Scanner | YES | YES | YES | YES |
Smart GFI | YES | YES | YES | YES |
Analog Meter | - | - | YES | YES |
RS232 Interface | Option | Option | YES | YES |
GPIB Interface | Option | Option | Option | Option |
Printer Interface | Option | Option | Option | Option |
TUV-GS listed | - | - | YES | YES |
CE Compliance | - | - | YES | YES |
Environment | 0-40℃, 20-80%RH | |||
Dimension (W×H×D) | 430×133×300 (mm) | 430×133×500( mm) | 430×133×500(mm) | |
Net Weight | 12Kg | 18Kg | 22Kg | |
7440STANDARD ACCESSORIES | ||||
Power Cord(10A) | ×1 | |||
Fuses | ×2 (Including a spare contained in the fuse holder) | |||
Hipot Test Lead ,1.5m(1101) | ×1 | |||
Return Test Cable(1102) | ×1 | |||
Hipot Return Lead ,1.8m(1103) | - | - | - | ×1 |
Ground Bond Return Lead 30A ,1.5m(1104) | - | - | - | ×1 |
Hipot Output Link Lead ,1.5m(1109) | - | ×8 | - | - |
RS232 Link Cable 1.5m(1130) | - | - | ×1 | ×1 |
Disk ( RS232 test program ) | - | - | ×1 | ×1 |
*product specifications are subject to change without notice. |